User Manual
2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-19
V-Source operating boundaries
Figure 6-11 and Figure 6-12 show the operating boundaries for the V-Source.
Only the first quadrant of operation is covered. Operation in the other three quad-
rants is similar.
Figure 6-11
V-Source output characteristics
210V21V
105mA
1.05A
A. Model 2400
1100V21V
21mA
1.05A
B. Model 2410
63V21V
1.05A
3.15A
C. Model 2420
21V
1.05A
3.15A
D. Models 2425 and 2430
10.5A
Model 2430
Pulse Mode
DC Mode
105V
Limit I Limit I
Limit I
Limit I
Source V
Source V
Source VSource V
42V10.5V
1.05A
5.25A
E. Model 2440
Limit I
Source V