User Manual

6-8 Source-Measure Concepts 2400 Series SourceMeter
®
User’s Manual
The measure time depends on the selected measurement speed. For example, if
speed is set at 0.01 PLC (power line cycles), the measure time would be 167µsec
for 60Hz operation (0.01/60).