User Manual

5-8 Pulse Mode Operation (Model 2430 only) 2400 Series SourceMeter
®
User’s Manual
NOTE With auto zero disabled, zero drift will eventually corrupt the accuracy of
the pulse measurement. To re-establish an accurate pulse measure
-
ment, enable auto zero or change the speed setting.
NOTE From the front panel, auto zero is controlled from the A/D CONTROLS
item of the Main Menu. For remote operation, the :SYSTem:AZERo com
-
mand controls auto zero.
Pulse-only
The fastest pulses are achieved by disabling measurements. With the signal not
measured, as shown in Figure 5-4, the pulse width can be as short as the
150µsec overhead. With reference and zero not measured, and display disabled,
the output off-time can be as short as the 1.9msec overhead.
NOTE Measurements cannot be disabled from the front panel. For remote oper-
ation, measurements can be disabled by sending the [SENSe]:FUNC-
tion:OFF:ALL command. See Section 18 for details. To disable the
display from the front panel, press CONFIG then EDIT, then select NOW.
Figure 5-4
Pulse-only timing
Pulse jitter
Pulse jitter is defined as the variance in the configured pulse on and off times. The
amount of jitter that can occur depends on how the SourceMeter is set up. For the
least amount of jitter (<5µsec for pulse on-time), turn off the display. See
Section 1, “Disabling front panel display.” With the display on, pulse on-time jitter
depends on the pulse width. For pulses <1msec wide, on-time jitter can be up to
20µsec. For pulses >1msec wide, on-time jitter is 5–40µsec.
PD
Delay = Pulse width delay, used to achieve pulse width setting.
150µs = Minimum pulse width overhead
1.90ms = Minimum output off-time overhead
PD = Pulse delay setting, used to determine time between pulses.
1.9ms
0V or 0A
Pulse Width
(Output On-Time)
Output Off-Time
Delay 150µs