User manual
    2260B Series User Manual 
80 
2.  Turn on the power. MS (Mass Storage) will be 
displayed on the screen after a few seconds if 
the USB drive is recognized. 
 Note 
If the USB drive is not recognized, check to see 
that the function settings for F-20 = 1 (page 90). If 
not, reinsert the USB flash drive. 
3.  Configure T-03 (Test Sequence 
Export) from 0~10 (internal 
memory slot) 
Page 76 
T-03 range 
1~10 
4.  The test sequence file will now be copied to the 
USB flash drive. 
 Note 
Error messages: If you try to export a test sequence 
from an empty memory location “Err 003” will be 
displayed on the display. 
Remove Test Sequence 
Overview 
The Remove Test Sequence function will delete 
a test sequence from the internal memory slot.  