Series Output 
Connection 
Master unit
Load
+
-
Slave unit
Steps 
1.  Ensure the power is off on both power supplies. 
2.  Choose a master and slave unit. 
3.  Connect the analog connectors for the master 
and slave unit as shown above. 
4.  Remove the output terminal cover 
and the protection dummy plug 
from the analog control connector. 
Page 36 
5.  Connect the master and slave unit in series as 
shown above. 
6.  Reattach the terminal cover. 
Page 36 
 Note 
Ensure load cables have sufficient 
current capacity. 
Page 33 
Re-attach the protection dummy plug when not in 
use. 
Master-Slave Series Operation 
Master-Slave 
Configuration 
Before using the power supplies in series, the 
master and slave units need to be configured. 
1.  Configure the OVP and OCP 
settings for the master unit. 
Page 43 