User manual
    2260B Series User Manual 
108 
4.  Press the Output key. The voltage 
can now be controlled with the 
External resistance. 
Note 
Ensure the resistor(s) and cables used exceed the 
isolation voltage of the power supply. For example: 
insulation tubes with a withstand voltage higher 
than the power supply can be used.  
When choosing an external resistor ensure the 
resistor can withstand a high degree of heat. 
Note 
CV and CC Slew Rate Priority are disabled for V-I 
mode (F-03) when using external resistance 
control. See the normal function settings on page 
87. 
External Resistance Control of Current Output 
Background 
External resistance control of the current output is 
accomplished using the MIL-26 connector on the 
rear panel. A resistance of 0kΩ~10kΩ is used to 
control the full scale current of the instrument. 
The output current (0 to full scale) can be 
controlled with the external resistance going up 
(Ext-R  ) 0kΩ~10kΩ(10kΩ = Vo,max) or down 
(Ext-R ) 10kΩ~0kΩ(10kΩ = 0). 
For 0kΩ~10kΩ: Output current = full scale current 
× (external resistance/10) 
For 10kΩ~0kΩ: Output current = full scale current 
× ([10-external resistance]/10) 