Datasheet

DC POWER SUPPLIES
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
Series 2200
Single-Channel Programmable
DC Power Supplies
The Series 2200 Single-Channel Programmable DC Power Supplies provide
a wide range of voltage outputs to address the testing and characteriza-
tion of components, circuits, modules, and complete devices, whether you
are in a research laboratory, in design and development, or in production
test. The Series 2200 consists of five models with output voltages from
20V to 72V that can deliver 86W, 96W, 100W, and 150W of power. In addi-
tion, these power supplies can act as constant current sources as well as
constant voltage sources. The Series 2200 power supplies offer an excellent
combination of performance, versatility, and ease of use that allow you to
obtain quality test data as quickly as possible. They perform as effectively
in automated test systems as they do in manual instrument configurations.
Outstanding Accuracy Delivered to the Load
With basic voltage setting accuracy of 0.03% and basic voltage readback
accuracy of 0.02%, you can be sure that the voltage you program for the
load is applied at the output terminals. What’s more, the rear panel con-
nections include remote sense terminals that compensate for voltage drops
in the power supply leads. This helps to ensure that the correct voltage is delivered to the load termi-
nals of the device-under-test (DUT). Great accuracy is not limited to voltage—the basic current setting
and readback accuracy is 0.05%, providing you with high quality load current measurements. Also,
with less than 5mVp-p noise, you can be confident that the power applied to the DUTs load termi-
nals is both accurate and of high quality.
Superior resolution is also provided by Keithley’s Series 2200 single-channel power supplies. With
1mV and 0.1mA resolution, the effects of very small changes in voltage and current can be detected
and studied. For portable devices in which minimum power consumption is critical, the 0.1mA cur-
rent resolution allows you to measure the idle and sleep mode currents so you can verify that your
products meet aggressive low power consumption goals.
Get Test Results Quickly
Keithleys Series 2200 single-channel power supplies have a number of features that enable you to
obtain the results you need quickly and easily, including tools to help you create sophisticated tests
for a wide range of requirements.
The dual-line display shows both the programmed settings and the actual voltage and current out-
puts, allowing you to immediately see, understand, and address any differences between the expect-
ed and actual output values. Multiple methods can be used to adjust the voltage and current settings.
You can use the direct-entry numeric keypad to set precise voltage and current values. There is also a
rotary knob with adjustable step size that lets you easily study the response of your DUT to small or
large changes in voltage or current.
Need to repeat a set of tests often? Instead of programming a number of parameters for each test
every time you run the test, just use a few keystrokes to save a test setup once and then recall it
whenever you need it. Take advantage of 40 memory locations to save up to 40 set ups or use the
Series 2200 List mode to define custom test sequences of up to 80 steps. This makes it easy to per-
form tests such as analyzing how your circuit- or device-under-test performs at each voltage level
within a range of voltages. A saved test can be run manually using front panel key strokes, automati-
cally using external trigger signals, or remotely using programmable interface commands. Up to
seven 80-step lists can be stored in a Series 2200 single-channel power supply. Each step can have a
programmable duration.
Protects Your DUT at All Times
A number of features are built into the Series 2200 power supplies to ensure that your DUT is pro-
tected from damage. A maximum voltage can be set so that regardless of the voltage value requested,
the output will not exceed the programmed limit value. For further voltage magnitude protection, an
Over Voltage protection level can be programmed that will cause the output to drop below 1V if the
Over Voltage limit is reached. These protections are in addition to the Current Limit setting, which
restricts the amount of current that can flow into the DUT. If the Current Limit is reached, the Series
Five models ranging in power
from 86W to 150W with voltage
outputs from 20V to 72V
address a wide range of power
requirements
0.03% basic voltage output
accuracy and 0.05% basic
current accuracy provide
quality test data
High output and measurement
resolution, 1mV and 0.1mA,
for testing low power circuits
and devices
Remote sensing to ensure the
programmed voltage is applied
to the load
Dual-line display shows both
the programmed values and
actual outputs for a continuous
indication of the status of the
power delivered to the load
Repeatable test sequences
of up to 80 output steps are
easy to create with the built-in
List mode
GPIB and USB interfaces
are standard for convenient
automated control
Single-channel programmable DC power supplies
Single-channel programmable DC power supplies