User manual
11-16 Remote Operation
Figure 11-5
Standard event status
Figure 11-6
Operation event status
* ESR ?
PON
(B7)
URQ
(B6)
CME
(B5)
EXE
(B4)
DDE
(B3)
QYE
(B2)
(B1) (B0)
OR
Standard Event
Status Register
Standard Event
Status Enable
Register
PON = Power On
URQ = User Request
CME = Command Error
EXE = Execution Error
DDE = Device-Dependent Error
QYE = Query Error
OPC = Operation Complete
&=Logical AND
OR = Logical OR
&
&
&
&
&
OPC
&
&
PON
(B7)
URQ
(B6)
CME
(B5)
EXE
(B4)
DDE
(B3)
QYE
(B2)
(B1) (B0)
OPC
* ESE
* ESE ?
To Event
Summary
Bit (ESB) of
Status Byte
Register (See
Figure 11-9).
(B15 - B8)
(B15 - B8)
OR
Idle = Idle state of the 2182
Filt = Filter Settled
Trig = Trigger Layer
Meas = Measuring
Cal = Calibrating
& = Logical AND
OR = Logical OR
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4)
(B3)
(B2)
(B1) (B0)
Operation Event
Register
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4)
(B3)
(B2)
(B1) (B0)
Operation Event
Enable Register
&
To Operation
Summary Bit
(OSB) of Status
Byte Register.
(See Figure 11-9).
&
&
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4)
(B3)
(B2)
(B1) (B0)
Operation
Condition Register
Filt
Filt
Cal
Cal
&
Filt
&
Cal