User manual
Stepping and Scanning 9-15
Set up 2182
Restore factory defaults
Filters: off
Rate: 1plc
Ch1: 10mV
Ch2: 1V
Ext Trigger: on
Delay: Set to time needed for cable settling
Config SCAN: INT
Timer off
Ch1 Count 3 ; Note Ch1 will store 3 readings / 2400 programmed
current level.
Ch2 will store 1 reading / 2400 programmed current level.
Rdg Count 48
Figure 9-5
Setup of Model 2182 and Model 2400
R
REF
2182
8501 Trigger Link Cable
2400
100Ω
GRN
Sample
WHT RED BLK
2107
Thermal
EMFs