Information

Model 2110 -Digit Digital Multimeter Instrument Specifications
Specifications are subject to change without notice
SPEC-2110 Rev. C / August 2013 3 of 4
AC CHARACTERISTICS: ACCURACY
6
± (% OF READING + % OF RANGE)
Function Range Resolution Frequency (Hz)
One year,
23 °C ±5 °C
Temperature coefficient
0 °C to 18 °C & 28 °C to 40 °C
Frequency
and period
100.000 mV
to 750.00 V
7
1 µV to
10 mV
10 to 40 0.03 0.002
40 to 300 k 0.02 0.002
AC T
RMS
voltage
100.000 mV
to 750.00 V
1 µV to
10 mV
10 to 20 k 0.12 + 0.05 0.01 + 0.01
20 K to 50 k 0.25 + 0.05 0.02 + 0.02
50 k to 100 k 0.65 + 0.08 0.04 + 0.02
100 k to 300 k 5.00 + 0.50 0.2 + 0.02
AC T
RMS
current
1.0000 A to
3.00000 A
10 µA to
100 µA
10 to 900 0.30 + 0.06 0.02 + 0.01
900 to 5 k 1.50 + 0.15 0.02 + 0.01
10.0000 A
100 µA
10 to 900 0.50 + 0.12 0.02 + 0.01
900 to 5 k 2.50 + 0.20 0.02 + 0.01
CAPACITANCE CHARACTERISTICS: ACCURACY
8
±
(% OF READING + % OF RANGE)
Range Test current One year, 23 °C ±5 °C
1.000 nF
10 µA
2.0 + 0.80
10.00 nF
10 µA
1.0 + 0.50
100.0 nF 100 µA 1.0 + 0.50
1.000 µF 100 µA 1.0 + 0.50
10.00 µF 100 µA 1.0 + 0.50
100.0 µF 1 mA 1.0 + 0.50
1000 µF 1 mA 1.0 + 0.50
10000 µF 1 mA 2.0 + 0.50
6
Slow AC filter (3 Hz bandwidth). Pure sine wave input greater than 5 % of range.
7
750 VAC range is limited to 100 kHz.
8
ADC set for continuous trigger operation. Null must be used.