Information
Model 2110 5½-Digit Digital Multimeter Instrument Specifications
Specifications are subject to change without notice
2 of 4 SPEC-2110 Rev. C / August 2013
MEASUREMENT NOISE REJECTION DC (60 HZ/50 HZ) AT 5½ DIGITS
CMRR
3
NMRR
4
120 dB 60 dB
3
For 1 kΩ unbalance in LO lead.
4
For line frequency ±0.1 %.
TEMPERATURE (THERMOCOUPLE) CHARACTERISTICS: ACCURACY
5
RTD AND NTC THERMISTOR MEASUREMENTS: ACCURACY ±0.8 °C, ONE YEAR, EXCLUSIVE OF LEAD
ACCURACY
PT100, D100, F100, PT385, PT3916, SPRTD (R-Zero, A4, B4, Ax, Bx, Cx, and Dx), NTCT (A, B, and C),
and user-definable RTD.
5
ADC set for continuous trigger operation
Thermocouple type Range ( °C) One year, exclusive of lead accuracy
B 600 to 1800 ±1.5 °C
C 0 to 2300 ±1.5 °C
E
−250 to 1000
±1.5 °C
J
−200 to 1200
±1.0 °C
K
−200 to 1350
±1.0 °C
N
−200 to 1300
±1.0 °C
R 0 to 1750 ±1.5 °C
S 0 to 1750 ±1.5 °C
T
−250 to 400
±1.5 °C