Information

Model 2110
5½-Digit Digital Multimeter
Instrument Specifications
Specifications are subject to change without notice
SPEC-2110 Rev. C / August 2013
*PSPEC-2110C*
1 of 4
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Model 2110 5½-Digit Digital
Multimeter instrument. Specifications are the standards against which the Model 2110 is tested. Upon
leaving the factory, the Model 2110 meets these specifications. Supplemental and typical values are
nonwarranted, apply at 23 °C, and are provided solely as useful information. Specifications are valid after a
two hour warm-up.
DC CHARACTERISTICS: ACCURACY
1
± (% OF READING + % OF RANGE)
Function Range Resolution
Input
resistance
One year,
23 °C ±5 °C
Temperature coefficient
0 °C to 18 °C & 28 °C to 40 °C
DC voltage
100.000 mV
1 µV
10 M
0.012 + 0.004
0.001 + 0.0005
1.00000 V
10 µV
0.012 + 0.001
0.0009 + 0.0005
10.0000 V 0.1 mV 0.012 + 0.002 0.0012 + 0.0005
100.000 V 1 mV 0.012 + 0.002 0.0012 + 0.0005
1000.00 V 10 mV 0.02 + 0.003 0.002 + 0.0015
DC current
10.0000 mA
0.1 µA 5.1
0.05 + 0.020
0.005 + 0.002
100.000 mA
1 µA 5.1
0.05 + 0.010
0.005 + 0.001
1.000000 A
10 µA 0.1
0.150 + 0.020
0.008 + 0.001
3.0000 A
100 µA 0.1
0.200 + 0.030
0.008 + 0.001
10.0000 A
100 µA 0.005
0.250 + 0.050
0.008 + 0.001
Resistance
2
100.000 1 m
1 mA 0.020 + 0.020
0.003 + 0.0005
1.00000 k 10 m
1 mA 0.020 + 0.003
0.003 + 0.0005
10.0000 k 100 m 100 µA
0.020 + 0.002
0.003 + 0.0005
100.000 k 1 10 µA
0.020 + 0.002
0.003 + 0.0005
1.00000 M 10 1 µA
0.030 + 0.004
0.005 + 0.0005
10.0000 M 100 0.1 µA
0.200 + 0.004
0.05 + 0.0005
100.000 M 1 k 0.1 µA
2.000 + 0.005
0.5 + 0.0005
Diode test
1.0000 V
10 µV
1 mA 0.020 + 0.030
0.002 + 0.0005
Continuity
1000 10 m
1 mA 0.020 + 0.020
0.002 + 0.0005
1
ADC set for continuous trigger operation. Input bias current <30 pA at 25 °C. Input protection 1000 V all ranges (2 wire input). Measurement
rate set to 10 PLC.
2
Specifications for 4-wire ohms mode. For 2-wire ohms, use zero null or subtract lead resistance from displayed reading. Maximum lead
resistance 10% of range per lead for 100 and 1 k ranges; add 1 k per lead for all other ranges.