3: Basic DMM operation Model 2110 5½ Digit Multimeter
Reference Manual
3-54 2110-901-01 Rev. C/August 2013
Remote operation
Commands that use the <n> parameter can use the MINimum and MAXimum name parameters. MIN
selects the minimum value and MAX selects the maximum value.
Use the following commands to configure and control limit testing:
CALCulate:FUNCtion LIMit Select the limits feature
CALCulate:LIMit:LOWer <n> Set the lower limit
CALCulate:LIMit:UPPer <n> Set the upper limit
<n> = (see Note) Set value (1 µ resolution)
The limit value setting range depends on the selected function. For example, for DCV, the limit value
can be set from -1200 to 1200.
CALCulate:STATe <b> Enable or disable math operations
<b> = 0 or OFF Disable math
= 1 or ON Enable math
mX+b
This math function multiplies a scale factor (m) to the measured reading (X), and then adds an offset
(b). The result of the calculation (y) is displayed on the instrument:
y = mX + b
You can enable this math function and set the values of m and b from the front panel or by using
remote operation. These values are stored in volatile memory and are lost when the instrument is
turned off.