User manual
Model 2110
5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-69
TRIGger:DELay:AUTO <b>
Turn automatic trigger delay on or off.
The delay is determined by function, range, integration time, and AC filter setting. Specifying a delay time
automatically turns off the automatic trigger delay.
Parameters
<b> description
0 or OFF
Turn selected automatic trigger delay off.
1 or ON
Turn automatic trigger delay on.
Query
command description
TRIGger:DELay:AUTO?
Query the automatic trigger delay mode. Returns 0 (OFF) or 1 (ON).
TRIGger:SOURce <name>
Use this command to select a trigger source. The instrument accepts a software (BUS) trigger, an immediate
internal trigger, or a hardware trigger from the rear-panel EXT TRIG terminal.
Parameters
<name> description
BUS
Select software trigger.
IMMediate
Select immediate internal trigger.
EXTernal
Select hardware trigger from the rear-panel EXT TRIG terminal.
Query
command description
TRIGger:SOURce?
Query the trigger source.
Details
Refer to Trigger sources (on page 3-30) for information about sources.