Specifications
DC CHARACTERISTICS
DC VOLTAGE
Accuracy
1
±(% of reading
+ % of range)
1 Year, 23° ±5°C
Temperature
Coefficient
18°C & 2–40°CRange Resolution
Input
Resistance
100.000 mV 1 µV
10 MW
0.012 + 0.004 0.001 + 0.0005
1.00000 V 10 µV 0.012 + 0.001 0.0009 + 0.0005
10.0000 V 0.1 mV 0.012 + 0.002 0.0012 + 0.0005
100.000 V 1 mV 0.012 + 0.002 0.0012 + 0.0005
1000.00 V 10 mV 0.02 + 0.003 0.002 + 0.0015
DCI (DC CURRENT)
Accuracy
1
±(% of reading
+ % of range)
1 Year, 23° ±5°C
Temperature
Coefficient
18°C & 2–40°CRange Resolution
Shunt
Resistance
10.0000 mA 0.1 µA
5.1 W
0.05 + 0.020 0.005 + 0.002
100.000 mA 1 µA
5.1 W
0.05 + 0.010 0.005 + 0.001
1.00000 A 10 µA
0.1 W
0.150 + 0.020 0.008 + 0.001
3.0000 A 100 µA
0.1 W
0.200 + 0.030 0.008 + 0.001
10.0000 A 100 µA
5 mW
0.250 + 0.050 0.008 + 0.001
RESISTANCE
2
Accuracy
1
±(% of reading
+ % of range)
1 Year, 23° ±5°C
Temperature
Coefficient
18°C & 2–40°CRange Resolution
Test
Current
100.000 W 1 mW
1 mA 0.020 + 0.020 0.003 + 0.0005
1.00000 kW 10 mW
1 mA 0.020 + 0.003 0.003 + 0.0005
10.0000 kW 100 mW
100 µA 0.020 + 0.002 0.003 + 0.0005
100.000 kW 1 W
10 µA 0.020 + 0.002 0.003 + 0.0005
1.00000 MW 10 W
1 µA 0.030 + 0.004 0.005 + 0.0005
10.0000 MW 100 W
0.1 µA 0.200 + 0.004 0.05 + 0.0005
100.000 MW 1 kW
0.1 µA 2.000 + 0.005 0.5 + 0.0005
DIODE TEST
Accuracy
1
±(% of reading
+ % of range)
1 Year, 23° ±5°C
Temperature
Coefficient
18°C & 2–40°CRange Resolution
Test
Current
1.0000V 10 µV 1 mA 0.020 + 0.030 0.002 + 0.0005
CONTINUITY
Accuracy
1
±(% of reading
+ % of range)
1 Year, 23° ±5°C
Temperature
Coefficient
18°C & 2–40°CRange Resolution
Test
Current
1000W 10 mW
1 mA 0.020 + 0.020 0.002 + 0.0005
1. Specifications valid after two hour warm-up.
a. ADC set for continuous trigger operation.
b. Input bias current <30pA at 25°C.
c. Measurement rate set to 10 PLC.
2. Specifications for 4W ohms mode. For 2W ohms, use zero null or subtract lead resistance from displayed reading.
a. Maximum lead resistance 10% of range per lead for 100W and 1kW ranges; add 1kW per lead for all
other ranges.
MEASUREMENT NOISE REJECTION DC (60Hz/50Hz) at 5.5 DIGITS
CMRR: 120dB for 1kW unbalance in LO lead.
NMRR: 60dB for line frequency ±0.1%.
TEMPERATURE (THERMOCOUPLE) CHARACTERISTICS
Thermocouple
Type Range
Accuracy
1
±°C
1 Year, exclusive of lead accuracy
B 600 to 1800°C 1.5
C 0 to 2300°C 1.5
E –250 to 1000°C 1.5
J –200 to 1200°C 1.0
K –200 to 1350°C 1.0
N –200 to 1300°C 1.0
R 0 to 1750°C 1.5
S 0 to 1750°C 1.5
T –250 to 400°C 1.5
1. Specifications valid after two hour warm-up;
a. ADC set for continuous trigger operation.
RTD AND NTC THERMISTOR MEASUREMENTS: Accuracy ±0.8˚C, 1 year, exclusive of lead
accuracy. PT100, D100, F100, PT385, PT3916, SPRTD (R-Zero, A4, B4, Ax, Bx, Cx, and Dx), NTCT
(A, B, and C), and user-definable RTD.
CAPACITANCE CHARACTERISTICS
Range
Test
Current
Accuracy
1
±(% of reading + % of range)
1 Year, 23° ±5°C
1.000 nF 10 µA 2.0 + 0.80
10.00 nF 10 µA 1.0 + 0.50
100.0 nF 100 µA 1.0 + 0.50
1.000 µF 100 µA 1.0 + 0.50
10.00 µF 100 µA 1.0 + 0.50
100.0 µF 1 mA 1.0 + 0.50
1. Specifications valid after two hour warm-up.
a. ADC set for continuous trigger operation.
b. Null enabled.
Model 2110 specifications
Model 2110 specifications
2110
5½-Digit Dual-Display Digital Multimeter
ACCESSORIES AVAILABLE
4299-3 Single Rack Mount Kit
4299-4 Dual Rack Mount Kit
4299-7 Fixed Rack Mount Kit
5805 Kelvin Probes, 0.9m (3ft)
5805-12 Kelvin Probes, 3.6m (12ft)
5808 Low Cost, Single Pin, Kelvin Probes
5809 Low Cost, Kelvin Clip Lead Set
6517-TP Thermocouple Bead Probe (K-Type)
7007-1 Shielded GPIB Cable, 1m (3.3 ft)
7007-2 Shielded GPIB Cable, 2m (6.6 ft)
8605 High Performance Modular Test Leads
8606 High Performance Modular Probe Kit
8680 RTD Probe Adapter
8681 Low Cost RTD
SERVICES AVAILABLE
2110-3Y-EW 1 Year Factory Warranty extended to 3 years from date of shipment
2110-5Y-EW 1 Year Factory Warranty extended to 5 years from date of shipment
C/2110-3Y-DATA 3 (Z-540-1 compliant) calibrations within 3 years of purchase for Model 2110
C/2110-5Y-DATA 5 (Z-540-1 compliant) calibrations within 5 years of purchase for Model 2110
C/2110-3Y-ISO 3 (ISO-17025 accredited) calibrations within 3 years of purchase for Model 2110
C/2110-5Y-ISO 5 (ISO-17025 accredited) calibrations within 5 years of purchase for Model 2110