Datasheet
DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
Voltage Range Resolution
Frequency
Range
Accuracy (1 Year)
23°C ±5°C
±(% of reading + % of range)
100 mV to 750 V 0.1 µV to 1 mV
3 Hz–10 Hz 0.35 + 0.03
10 Hz–20 kHz 0.06 + 0.03
20 kHz–50 kHz 0.12 + 0.05
50 kHz–100 kHz 0.60 + 0.08
100 kHz–300 kHz 4 + 0.5
AC OPERATING CHARACTERISTICS
Function Digits Readings/s Rate Bandwidth
ACV (all ranges), and
ACI (all ranges)
6½ 0.5 (0.4) SLOW 3 Hz–300 kHz
6½ 1.4 (1.5) MED 30 Hz–300 kHz
6½ 4.0 (4.3) MED 30 Hz–300 kHz
6½ 2.2 (2.3) FAST 300 Hz–300 kHz
6½ 35 (30) FAST 300 Hz–300 kHz
FREQUENCY AND PERIOD CHARACTERISTICS
ACV
Range
Frequency
Range
Period
Range
Gate
Time
Resolution
±(ppm of
reading)
Accuracy
90 Day/1 Year
±(% of reading)
100 mV to
750 V
3 Hz to
500 kHz
333 ms to
2 µs
1 s 0.3 0.01
TEMPERATURE CHARACTERISTICS
Thermocouple Accuracy
1
90 Day/1 Year (23°C ± 5°C)
Type Range Resolution
Relative to
Reference Junction
USING
2001-TCSCAN
2
J –200 to + 760°C 0.001°C ±0.5°C ±0.65°C
K –200 to + 1372°C 0.001°C ±0.5°C ±0.70°C
N –200 to + 1300°C 0.001°C ±0.5°C ±0.70°C
T –200 to + 400°C 0.001°C ±0.5°C ±0.68°C
4-WIRE RTD Accuracy
3
90 Day/1 Year
(23°C ± 5°C)
Accuracy
3
2 Years
(23°C ± 5°C)
Range Resolution
–100° to +100°C 0.001°C ±0.08°C ±0.12°C
–200° to +630°C 0.001°C ±0.14°C ±0.18°C
TEMPERATURE NOTES
1. For temperatures <–100°C, add ±0.1°C and >900°C add ±0.3°C.
2. Specifications apply to channels 2–6. Add 0.06°C/channel from channel 6.
3. Excluding probe errors.
2010
Low Noise 7½-Digit
Autoranging Multimeter
GENERAL
POWER SUPPLy: 100V / 120V / 220V / 240V.
LINE FREQUENCy: 50Hz to 60Hz and 440Hz, automatically sensed at power-up.
POWER CONSUMPTION: 22VA.
VOLT HERTz PRODUCT: ≤8 × 10
7
V·Hz.
OPERATING ENVIRONMENT: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C.
STORAGE ENVIRONMENT: –40° to 70°C.
ALTITUDE: Up to 2000 meters.
SAFET y: Conforms to European Union Directive 73/23/EEC EN 61010-1, Cat II.
EMC: Complies with European Union Directive 89/336/EEC, EN 61326-1.
VIBRATION: MIL-PRF-28800F Class 3 Random.
WARMUP: 2 hours to rated accuracy.
DIMENSIONS:
Rack Mounting: 89mm high × 213mm wide × 370mm deep (3½ in × 8
3
⁄8 in × 14
9
⁄16 in).
Bench Configuration (with handle and feet): 104mm high × 238mm wide × 370mm
deep (4
1
⁄8 in × 9
3
⁄8 in × 14
9
⁄16 in).
SHIPPING WEIGHT: 5kg (11 lbs).
DC VOLTAGE
Accuracy 23°C ± 5°C
±(ppm of rdg. +
ppm of range)
Input
ResistanceRange Resolution 90 Day 1 Year
100.00000 mV 10 nV 25 + 9 37 + 9
> 10 GW
1.0000000 V 100 nV 18 + 2 25 + 2
> 10 GW
10.000000 V 1 µV 18 + 4 24 + 4
> 10 GW
100.00000 V 10 µV 25 + 5 35 + 5
10 MW ±1%
1000.0000 V 100 µV 31 + 6 41 + 6
10 MW ±1%
RESISTANCE
Accuracy 23°C ± 5°C
±(ppm of rdg. +
ppm of range)
Test CurrentRange Resolution 90 Day 1 Year
10.000000 W 1 µW
40 + 9 60 + 9 10 mA
100.00000 W 10 µW
36 + 9 52 + 9 1 mA
1.0000000 kW 100 µW
33 + 2 50 + 2 1 mA
10.000000 kW 1 mW
32 + 2 50 + 2 100 µA
100.00000 kW 10 mW
40 + 4 70 + 4 10 µA
1.0000000 MW 100 mW
50 + 4 70 + 4 10 µA
10.000000 MW 1 W
200 + 4 400 + 4
640 nA//10 MW
100.00000 MW 10 W
1500 + 4 1500 + 4
640 nA//10 MW
DC CURRENT
Accuracy 23°C ± 5°C
±(ppm of rdg. +
ppm of range)
Burden
VoltageRange Resolution 90 Day 1 Year
10.000000 mA 1 nA 300 + 80 500 + 80 < 0.15 V
100.00000 mA 10 nA 300 + 800 500 + 800 < 0.18 V
1.0000000 A 100 nA 500 + 80 800 + 80 < 0.35 V
3.000000 A 1 µA 1200 + 40 1200 + 40 < 1 V
CONTINUITY 2W
Accuracy 23°C ± 5°C
±(ppm of rdg. +
ppm of range)
Test CurrentRange Resolution 90 Day 1 Year
1 kW 100 mW
100 + 100 120 + 100 1 mA
DIODE TEST
Accuracy 23°C ± 5°C
±(ppm of rdg. +
ppm of range)
Test CurrentRange Resolution 90 Day 1 Year
10.000000 V 1 µV 30 + 7 40 + 7 1 mA
4.400000 V 1 µV 30 + 7 40 + 7 100 µA
10.000000 V 1 µV 30 + 7 40 + 7 10 µA
DC OPERATING CHARACTERISTICS
Function Digits Readings/s PLCs
DCV (all ranges),
DCI (all ranges), and
Ohms (<10M range)
7½ 4 (3) 5
6½ 30 (27) 1
6½ 50 (44) 1
5½ 260 (220) 0.1
5½ 490 (440) 0.1
5½ 1000 (1000) 0.04
4½ 2000 (1800) 0.01
DC NOISE PERFORMANCE
Rate Digits
RMS Noise
100mV
Range
(2 min.)
RMS Noise
10V
Range
(2 min.) NM
RR CMRR
5 PLC 7½ 110 nV 1.2 µV 60 dB 140 dB
1 PLC 6½ 125 nV 1.4 µV 60 dB 140 dB
0.1 PLC 5½ 1.9 µV 11.5 µV — 80 dB
0.01 PLC 4½ 2.9 µV 139 µV — 80 dB
Model 2010 specifications
Model 2010 specifications