User manual

SCPI Conformance Information
2001-900-01 Rev. K / August 2010 I-5
[:SENSe[1]]
:RESistance
:APERture <n>
:APERture?
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:REFerence?
:OCOMpensated <b>
:OCOMpensated?
Path to configure resistance:
Specify integration rate in seconds (166.67e-6 to 200e-3).
Query aperture (integration rate).
Specify integration rate (number of line cycles; 0.01 to 10).
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to +1.05e9).
Query range.
Enable (1 or ON) or disable (0 or OFF) auto range.
Set range based on present input signal.
Query auto range.
Specify reference (REL) value (-1.05e9 to +1.05e9).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Query reference (REL) value.
Enable (1 or ON) or disable (0 or OFF) Offset Compensation.
Query Offset Compensation.
[:SENSe[1]]
:FRESistance
:APERture <n>
:APERture?
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:REFerence?
:OCOMpensated <b>
:OCOMpensated?
Path to configure four-wire resistance:
Specify integration rate in seconds (166.67e-6 to 200e-3).
Query aperture (integration rate).
Specify integration rate (number of line cycles; 0.01 to 10).
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to +2.1e5).
Query range.
Enable (1 or ON) or disable (0 or OFF) auto range.
Set range based on present input signal.
Query auto range.
Specify reference (REL) value (-2.1e5 to +2.1e5).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Query reference (REL) value.
Enable (1 or ON) or disable (0 or OFF) Offset Compensation.
Query Offset Compensation.
Command Description