User manual
SCPI Conformance Information
2001-900-01 Rev. K / August 2010 I-3
:ROUTe
:CLOSe <list>
:STATe?
:CLOSe? <list>
:OPEN <list>
:OPEN:ALL
:OPEN? <list>
:SCAN
[:INTernal] <list>
[:INTernal]?
Path and command to close specified channel:
Query closed channel.
Query specified channels (1 = closed, 0 = open).
Open specified channel or all channels.
Open all channels.
Query specified channels (1 = open, 0 = closed).
Command path to configure and control scan operations:
Specify an internal scan list (2 to 10 channels).
Query the internal scan list.
[:SENSe[1]]
:FUNCtion <name>
:FUNCtion?
:DATA?
Select measurement function: ‘VOLTage:AC’, ‘VOLTage:DC’, RESis-
tance’,‘CURRent:AC’, ‘CURRent:DC’, ‘FREQuency’, TEMPerature’
Query function.
Query latest reading.
[:SENSe[1]]
:CURRent:AC
:APERture <n>
:APERture?
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:REFerence?
Path to configure AC current:
Specify integration rate in seconds (166.67e-6 to 200e-3).
Query aperture (integration rate).
Specify integration rate (number of line cycles; 0.01 to 10).
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to +2.1).
Query range.
Enable (1 or ON) or disable (0 or OFF) auto range.
Set range based on present input signal.
Query auto range.
Specify reference (REL) value (-2.1 to +2.1).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Query reference (REL) value.
[:SENSe[1]]
:CURRent:DC
:APERture <n>
:APERture?
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:REFerence?
Path to configure DC current.
Specify integration rate in seconds (166.67e-6 to 200e-3).
Query aperture (integration rate).
Specify integration rate (number of line cycles; 0.01 to 10).
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to +2.1).
Query range.
Enable (1 or ON) or disable (0 or OFF) auto range.
Set range based on present input signal.
Query auto range.
Specify reference (REL) value (-2.1 to +2.1).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Query reference (REL) value.
Command Description