I.1 Introduction
The Model 2001 complies with SCPI version 1991.0. Table I-3 lists the SCPI confirmed commands implemented by the Model
2001, and Table I-4 lists the non-SCPI commands implemented.
q~ÄäÉ=fJP=póåí~ñ=çÑ=p`mf=ÅçåÑáêãÉÇ=Åçãã~åÇë=áãéäÉãÉåíÉÇ=Äó=íÜÉ=jçÇÉä=OMMN
Command Description
:CALCulate[1]
:FORMat <name>
:FORMat?
:STATe <b>
:STATe?
:DATA?
:IMMediate
Subsystem to control CALC 1:
Select math format: MXB, PERCent, NONE.
Query math format.
Enable (1 or ON) or disable (0 or OFF) calculation.
Query state of math function.
Read math result of CALC 1.
Recalculate input data.
:CALCulate2
:FORMat <name>
:FORMat?
:STATe <b>
:STATe?
:IMMediate
:DATA?
Subsystem to control CALC 2.
Select math format: MEAN, SDEViation, MAXimum, MINimum, PKPK,
NONE.
Query math format.
Enable (1 or ON) or disable (0 or OFF) calculation.
Query state of math function.
Recalculate raw input data in buffer.
Read math result of CALC 2.
:CALCulate3
:LIMit[1]
:UPPer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
:LOWer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
Subsystem to control CALC 3 (limit tests):
Path to control LIMIT 1 test:
Path to configure upper limit:
Specify upper limit (-9.99e35 to +9.99e35).
Query upper limit.
Specify digital output (0 to 15).
Query source.
Path to configure lower limit:
Specify lower limit (-9.99e35 to +9.99e35).
Query lower limit.
Specify digital output (0 to 15).
Query source.