User manual

IEEE-488 Conformance Information
2001-900-01 Rev. K / August 2010 H-3
q~ÄäÉ=eJOW=
Coupled commands
Command Also changes To If
:SYST:AMET BURS
:SYST:AMET:ASTR
and
:SYST:AMET:SSTR
:TRIG:COUN
:TRAC:FEED:CONT
... :RANG:AUTO
... :NPLC
... :APER
... :DIG:AUTO
... :DIG
... :AVER:AUTO
... :AVER:TCON
:INIT:CONT
:SYST:AZER:STAT
:ARM:LAY1:COUN
:ARM:LAY2:COUN
:TRIG:COUN
:TRACe:EGR
:TRACe:FEED
:TRACe:FEED:CONT
:SYST:AZER:STAT
:SYST:LSYN:STAT
:CALC:STAT
... :RANG:AUTO
... :NPLC
... :NPLC:AUTO
... :DIG
... :DIG:AUTO
... :AVER:STAT
... :AVER:AUTO
... :REF:STAT
:TRAC:POIN:AUTO
:TRAC:POIN
:TRAC:FEED:PRET:AMO:READ
:TRIG:SOUR
OFF
0.01
0.01 / line-frequency
OFF
4
OFF
MOV
OFF
OFF
1
1
TRACe:POIN
COMP
CALC1
NEXT
OFF
OFF
OFF
OFF
0.01
OFF
4
OFF
OFF
OFF
OFF
OFF
TRIG:COUN
TRIG:COUN *
TRAC:FEED:PRET:AMO:
PERC / 100
TIM
TRACe:FEED is NONE
TRAC:POIN:AUTO is ON and
TRIG:COUN is an illegal buffer
size
TRAC:POIN:AUTO is ON
TRAC:POIN:AUTO is ON
TRAC:POIN:AUTO is ON
TRAC:FEED:CONT is PRET
and TRIG:SOUR is neither