User manual

H
IEEE-488 Conformance Information
H.1 Information
The IEEE-488.2 standard requires specific information about how the Model 2001 implements the standard. See paragraph 4.9
of the IEEE-488.2 standard (Std 488.2-1987) for a list of the documentation requirements. Table H-1 provides a summary of
the requirements, and then provides the information or references the manual for that information.
q~ÄäÉ=eJNW=
SCPI conformance
Requirement Description or reference
(1)
(2)
(3)
(4)
IEEE-488 Interface Function Codes
Behavior of 2001 when the address is set outside the range 0-
30
Behavior of 2001 when valid address is entered
Power-On Setup Conditions
See Appendix D
Cannot enter an invalid address
Address changes and bus resets
Determined by :POSetup <name>
(5)
(a)
(b)
(c)
(d)
(e)
Message Exchange Options:
Input buffer size.
Queries that return more than one response message unit
Queries that generate a response when parsed
Queries that generate a response when read
Coupled commands
256 bytes
None
All queries (Common Commands and SCPI)
None
See Table H-2
(6) Functional elements required for SCPI commands Contained in Appendix C
(7) Buffer size limitations for block data Block display messages: See paragraph 4.12 through
paragraph 4.25