User manual

IEEE-488 Reference
2001-900-01 Rev. K / August 2010 4-175
Programming example 10 OUTPUT 716; “:stat:oper:enab 514; enab?”
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10 Sets Bits B1 and B9 of the Operation Event Enable Register, and then queries the register.
Line 20 Addresses the Model 2001 to talk.
Line 30 Displays the decimal value that defines which bits in the register are set (514).
4.21.3 :PTRansition <NRf>
:STATus:MEASurement:PTRansition <NRf> Program Measurement Transition Filter (PTR)
:STATus:QUEStionable:PTRansition <NRf> Program Questionable Transition Filter (PTR)
:STATus:OPERation:PTRansition <NRf> Program Operation Transition Filter (PTR)
:STATus:OPERation:TRIGger:PTRansition <NRf>Program Trigger Transition Filter (PTR)
:STATus:OPERation:ARM:PTRansition <NRf>Program Arm Transition Filter (PTR)
:STATus:OPERation:ARM:SEQuence:PTRansition <NRf>Program Sequence Transition Filter (PTR)
Parameters <NRf> = 0 Clear PTR register
= 1 Set bit B0 of PTR registe
= 2 Set bit B1 of PTR register
= 4 Set bit B2 of PTR register
= 8 Set bit B3 of PTR register
= 16 Set bit B4 of PTR register
= 32 Set bit B5 of PTR register
= 64 Set bit B6 of PTR register
= 128 Set bit B7 of PTR register
= 256 Set bit B8 of PTR register
= 512 Set bit B9 of PTR register
= 1024 Set bit B10 of PTR register
= 2048 Set bit B11 of PTR register
= 16384 Set bit B14 of PTR register
Formats :stat:meas:ptr <NRf>
:stat:ques:ptr <NRf>
:stat:oper:ptr <NRf>
:stat:oper:trig:ptr <NRf>
:stat:oper:arm:ptr <NRf>
:stat:oper:arm:seq:ptr <NRf>
Defaults Power-up Sets (1) all bits
*CLS No effect