User manual
IEEE-488 Reference
2001-900-01 Rev. K / August 2010 4-169
Trigger Event Register:
Not used.
Set bit indicates that the instrument is in the trigger layer (PTR), or that
the instrument has exited from the trigger layer (NTR).
Not used.
Arm Event Register:
Not used.
Set bit indicates that the instrument is in an arm layer (PTR), or that the
instrument has exited from the arm layers (NTR).
Not used.
Idle Arm Trig
B15 - B11 B10
B9 B8 B7 B6 B5
B4 B3
B2 B1
B0
(2 )
10
1024
(2 )
9
(2 )
6
(2 )
5
(2 )
1
(2 )
0
64 32
0/1 0/1 0/1
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Operation Event Set
0 = Operation Event Cleared
Events : Idle = Idle state of the 2001
Calc = Calculating
Arm = Waiting for Arm
Trig = Waiting for Trigger
Set = Settling
Cal = Calibrating
Set
2
Calc Cal
512 1
0/1 0/1
cáÖìêÉ=QJOQ
Operation Event Register
cáÖìêÉ=QJOR
Trigger Event Register
B15 - B2
B1 B0
(2 )
1
2
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Event Bit Set
0 = Event Bit Cleared
Event : Seq1 = Sequence 1
Seq1