User manual

IEEE-488 Reference
4-168 2001-900-01 Rev. K / August 2010
Operation Event Register:
Set bit indicates that the instrument is calibrating (PTR), or cali-
bration is complete (NTR).
Set bit indicates that the instrument is waiting for signals to settle
(PTR), or the settling period has expired (NTR).
Not used.
Set bit indicates that the instrument is waiting in the trig-
ger layer (PTR) of the trigger model, or that it has left the trigger layer (NTR).
Set bit indicates that the instrument is waiting in an arm
layer of the trigger model (PTR), or operation has proceeded from the arm layers into the trigger
layer (NTR).
Not used.
Set bit indicates that the instrument is performing a math calcu-
lation (PTR) or the calculation is complete (NTR).
Set bit indicates that the instrument is in the idle state (PTR), or has left the
idle state to perform an operation (NTR).
Not used.
Cal
B13 - B9 B8
B15 B14
(2 )
8
256
(2 )
14
16384
0/10/1
Warn
B7 - B5
0
Temp
B4
(2 )
4
16
0/1
B3 - B0
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Event Bit Set
0 = Event Bit Cleared
Events : Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
cáÖìêÉ=QJOP
Questionable Event Register