IEEE-488 Reference
2001-900-01 Rev. K / August 2010 4-159
Description This command is used to specify the simulated reference temperature. The temperature value depends on
which temperature scale is currently selected (°C, °F or K). Typically, or 23°C is used as the simulated
reference temperature.
The :ACQuire command is then used to store the simulated reference temperature.
Programming example 10 OUTPUT 716; “:unit:temp c”
20 OUTPUT 716; “:temp:rjun1:sim 0; sim?”
30 ENTER 716; A$
40 PRINT A$
50 END
Line 10 Selects °C temperature scale.
Line 20 Two commands in this program message; the first programs reference junction #1 for a 0°C
simulated reference temperature, and the second queries the simulated reference temperature.
Line 30 Addresses the Model 2001 to talk.
Line 40 Displays the simulated reference temperature (0)
4.19.23 :REAL commands
:TCOefficient <n>
[:SENSe[1]]:TEMPerature:RJUNctionX:REAL:TCOefficient <n>Specify temperature coefficient
Parameters <n> = -0.09999 to +0.09999Specify temperature coefficient
= DEFault +0.01 temperature coefficient
= MINimum -0.09999 temperature coefficient
= MAXimum +0.09999 temperature coefficient
Format :temp:rjunX:real:tco <n>
where; X = 1 to 5 (reference junction #)
Defaults Power-up Saved power-on setup
*RST 0.01 temperature coefficient
:SYSTem:PRESet 0.01 temperature coefficient
Query :TCOefficient? Query temperature coefficient (TC)
:TCOefficient? DEFaultQuery *RST default TC
:TCOefficient? MINimumQuery lowest allowable TC
:TCOefficient? MAXimumQuery largest allowable TC
Short-form format: :temp:rjunX:real:tco?
:temp:rjunX:real:tco? def
:temp:rjunX:real:tco? min
:temp:rjunX:real:tco? max
where; X = 1 to 5 (reference junction #)
Response message: -0.09999 to +0.09999
Description This command is used to specify the temperature coefficient (TC) of the “real” temperature reference
junction. TC is specified in °C/volt and is not affected by the :UNIT :TEMPerature command. The :AC-
Quire command can then be used to acquire the reference temperature.