User manual
IEEE-488 Reference
4-136 2001-900-01 Rev. K / August 2010
q~ÄäÉ=QJOO
Auto resolution; DCI
Measurement type NPLC* selection
Resolution
(Auto resolution enabled)
Normal
Normal
Normal
Normal
Normal
In-circuit
2 to 10**
0.2 to <2**
0.02 to <0.2**
0.01 to <0.02**
Auto NPLC or auto aperture enabled
Any
7H digits
6H digits
5H digits
4H digits
6H digits
5H digits
* To convert NPLC values to aperture values, see :APERture.
** Auto NPLC and auto aperture disabled.
q~ÄäÉ=QJOP
Auto resolution; ACV
Measurement type NPLC* selection
Resolution
(Auto resolution enabled)
RMS, average
RMS, average
RMS, average
RMS, average
Low frequency RMS
Peak, positive and negative peak spikes
10**
0.02 to <10**
0.01 to <0.02**
Auto NPLC or auto aperture enabled
Any
Any
6H digits
5H digits
4H digits
5H digits
5H digits
3H digits
* To convert NPLC values to aperture values, see :APERture.
** Auto NPLC and auto aperture disabled.
q~ÄäÉ=QJOQ
Auto resolution; DCV,
2 and
4
NPLC* selection
Resolution
(Auto resolution enabled)
2 to 10**
0.2 to <2**
0.02 to <0.2**
0.01 to <0.02**
Auto NPLC or auto aperture enabled
7H digits
6H digits
5H digits
4H digits
6H digits
* To convert NPLC values to aperture values, see :APERture.
** Auto NPLC and auto aperture disabled.