User manual
IEEE-488 Reference
2001-900-01 Rev. K / August 2010 4-127
:ULIMit <n>
[:SENSe[1]]:CURRent:AC:RANGe:AUTO:ULIMit <n>Set upper limit for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO:ULIMit <n>Set upper limit for DCI
[:SENSe[1]]:VOLTage:AC:RANGe:AUTO:ULIMit <n>Set upper limit for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO:ULIMit <n>Set upper limit for DCV
[:SENSe[1]]:RESistance:RANGe:AUTO:ULIMit <n>Set upper limit for 2
[:SENSe[1]]:FRESistance:RANGe:AUTO:ULIMit <n>Set upper limit for 4
:LLIMit <n>
[:SENSe[1]]:CURRent:AC:RANGe:AUTO:LLIMit <n>Set lower limit for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO:LLIMit <n>Set lower limit for DCI
[:SENSe[1]]:VOLTage:AC:RANGe:AUTO:LLIMit <n>Set lower limit for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO:LLIMit <n>Set lower limit for DCV
[:SENSe[1]]:RESistance:RANGe:AUTO:LLIMit <n>Set lower limit for 2
[:SENSe[1]:FRESistance:RANGe:AUTO:LLIMit <n>Set lower limit for 4
Parameters <n> = 0 to 2.1 Range limit for ACI and DCI
= 0 to 775 Range limit for ACV
= 0 to 1100 Range limit for DCV
= 0 to 1.05e9 Range limit for 2
= 0 to 2.1e5 Range limit for 4
= DEFault 2.1 (ACI and DCI)
775 (ACV)
1100 (DCV)
1.05e9 (2)
2.1e5 (4)
= MINimum 0 (All functions)
= MAXimum Same as DEFault
Formats :curr:ac:rang:auto:ulim <n>
:curr:ac:rang:auto:llim <n>
:curr[:dc]:rang:auto:ulim <n>
:curr[:dc]:rang:auto:llim <n>
:volt:ac:rang:auto:ulim <n>
:volt:ac:rang:auto:llim <n>
:volt[:dc]:rang:auto:ulim <n>
:volt[:dc]:rang:auto:llim <n>
:res:rang:auto:ulim <n>
:res:rang:auto:llim <n>
:fres:rang:auto:ulim <n>
:fres:rang:auto:llim <n>
Defaults Power-up Saved power-on setup
*RST Upper limit Lower limit
ACI and DCI 2.1 0.0002
ACV 775 0.2
DCV 1100 0.2