User manual

IEEE-488 Reference
2001-900-01 Rev. K / August 2010 4-123
Programming example 10 OUTPUT 716; “:curr:ac:nplc:auto on; auto?”
20 ENTER 716; A$
30 PRINT A
40 END
Line 10 Enables auto NPLC and then queries the state of auto NPLC.
Line 20 Addresses the Model 2001 to talk.
Line 30 Displays the state of auto NPLC (1).
4.19.7 :COUPling AC|DC
[:SENSe[1]]:CURRent:AC:COUPling AC|DC Specify coupling for ACI
[:SENSe[1]]:VOLTage:AC:COUPling AC|DC Specify coupling for ACV
[:SENSe[1]]:FREQuency:COUPling AC|DC Specify coupling for FREQ
Parameters AC AC coupling
DC AC+DC coupling
Formats :curr:ac:coup ac :volt:ac:coup ac :freq:coup ac
:curr:ac:coup dc:volt:ac:coup dc:freq:coup dc
Defaults Power-up Saved power-on setup
*RST AC
:SYSTem:PRESet AC
Query :COUPling? Query currently programmed coupling
Short-form format: :curr:ac:coup?
:volt:ac:coup?
:freq:coup?
Response message: AC or DC
Description This command is used to select the input coupling for the AC, ACV and the FREQ functions. When AC
coupling is selected, a DC blocking capacitor is placed in series with the AC measurement circuit. This
removes the DC component from the measurement.
q~ÄäÉ=QJOM
Auto NPLC; DCV,
2 and
4
Resolution
NPLC value
(Auto NPLC enabled)
7H digits*
6H digits*
5H digits*
4H digits*
3H digits*
Auto resolution enabled
2
0.2
0.02
0.01
0.01
1
* Auto resolution disabled.