Line 10Sets aperture for ACI and then queries the programmed aperture.
Line 20Addresses the Model 2001 to talk.
Line 30Displays the programmed aperture value.
:AUTO <b>|ONCE
[:SENSe[1]]:CURRent:AC:APERture:AUTO <b>|ONCEControl auto aperture for ACI
[:SENSe[1]]:CURRent[:DC]:APERture:AUTO <b>|ONCEControl auto aperture for DCI
[:SENSe[1]]:VOLTage:AC:APERture:AUTO <b>|ONCEControl auto aperture for ACV
[:SENSe[1]]:VOLTage[:DC]:APERture:AUTO <b>|ONCEControl auto aperture for DCV
[:SENSe[1]]:RESistance:APERture:AUTO <b>|ONCEControl auto aperture for 2
[:SENSe[1]]:FRESistance:APERture:AUTO <b>|ONCEControl auto aperture for 4
[:SENSe[1]]:TEMPerature:APERture:AUTO <b>|ONCEControl auto aperture for TEMP
<b>= 1 or ON Enable auto aperture
= 0 or OFFDisable auto aperture
ONCEEnable and then disable auto aperture
:curr:ac:aper:auto <b>:curr[:dc]:aper:auto <b>
:curr:ac:aper:auto ONCE:curr[:dc]:aper:auto ONCE
:volt:ac:aper:auto <b>:volt[:dc]:aper:auto <b>
:volt:ac:aper:auto ONCE:volt[:dc]:aper:auto ONCE
:res:aper:auto <b>:fres:aper:auto <b>
:res:aper:auto ONCE:fres:aper:auto ONCE
:temp:aper:auto <b>
:temp:aper:auto ONCE
Power-upSaved power-on setup
*RSTOFF
:SYSTem:PRESetOFF
:AUTO?Query state of auto aperture.
Short-form formats::curr:ac:aper:auto?