User manual

IEEE-488 Reference
4-54 2001-900-01 Rev. K / August 2010
:CONFigure:<function>
where; <function> = VOLTage[:DC] DCV function
= CURRent[:DC] DCI function
= VOLTage:AC ACV function
= CURRent:AC ACI function
= RESistance 2 function
= FRESistance 4 function
= TEMPerature TEMP function
= FREQuency FREQ function
= VOLTage:FREQuency FREQ function (voltage input)
= CURRent:FREQuency FREQ function (current input)
Formats :conf:volt[:dc]
:conf:curr[:dc]
:conf:volt:ac
:conf:curr:ac
:conf:res
:conf:fres
:conf:temp
:conf:freq
:conf:volt:freq
:conf:curr:freq
Query :CONFigure? Query the selected function
Short-form format: :conf?
Response message: CURR:DC, VOLT:AC, CURR:AC, VOLT:DC, RES, FRES, TEMP,
FREQ, VOLT:FREQ or CURR:FREQ
Description This command is used to configure the instrument for subsequent measurements on the specified
function. Basically, this command places the instrument in a “one-shot” measurement mode.
The :READ? command can then be used to trigger a measurement and acquire a reading (see
:READD?).
Note that there are three <function> versions to configure frequency; FREQuency,
VOLTage:FREQuency and CURRent:FREQuency. When FREQuency is sent, the *RST input
source is selected (voltage). When :VOLTage:FREQuency is sent, voltage input is selected.
When CURRent:FREQuency is sent, current input is selected.
When this command is sent, the Model 2001 will be configured as follows:
•The function specified by this command is selected.
•All controls related to the selected function are defaulted to the *RST values.
•Continuous initiation is disabled (:INITiate:CONTinuous OFF).
•The control sources in all three layers of the Trigger Model are set to Immediate.
•The count values in all three layers of the Trigger Model are set to one (1).
•The delay in all three layers of the Trigger Model are set to zero (0).
•The Model 2001 is placed in the idle state.