User manual
Getting Started
2-6 2001-900-01 Rev. K / August 2010
Buffered readings display
After readings have been stored in the buffer, they are
displayed on the front panel by pressing the RECALL key.
This action brings up the first reading in the buffer, such as:
+00.00000 VDC
Rdg#+00000 @Time=+000.000000 sec
where the top line shows the reading, and the bottom line
shows the reading number and time-stamp. The RANGE
and keys scroll through the readings and the and
keys change the digit being scrolled. You can scroll through
statistical data on the buffered readings by pressing the
NEXT and PREVious DISPLAY keys. This also gives you
the option of printing the data.
Buffered readings are preserved across
power transition when the memory
options are installed.
Message displays
While operating the Model 2001, the front panel display is
also used for showing status and error messages. These
messages are shown to inform you of parameter conflicts,
trigger overruns, etc. Refer to paragraph 3.3.4 for a
complete list of these messages.
Section 4, IEEE-488 Reference, describes the programming
of user messages to be shown on the front panel. (See
:DISPlay subsystem.) Both lines of the display are available
for this purpose.
2.3 Overview of measurement process
The following overview is intended to acquaint you with the
basic measurement fundamentals without overwhelming
you with the details of enhanced capabilities. This brief
overview is sufficient to support the operation examples in
paragraph 2.5 (front panel) and paragraph 2.6 (IEEE-488
bus). For a complete explanation on all aspects of the
measurement process, see paragraph 3.7.
The simplified model for a measurement operation is shown
in Figure 2-3. As shown, the measurement operationconsists
of three layers: the arm layer, scan layer, and measure layer.
q~ÄäÉ=OJO
Menu summary
CONFIG-DCV Press the CONFIG key, then the DCV
key, to view the top level of the DCV
configuration menu. Access to other
function and operation configurations are
similar, e.g., CONFIG-ACV or CON-
FIG-TRIG.
MENU Press the MENU key to view the top
level of the main menu. The operations
that have no corresponding key are
included in the main menu.
or Use the cursor keys to move the high-
lighted cursor among menu selections, or
the digits of a parameter value, or change
channels on the scanner.
RANGE
RANGE
Use the RANGE keys to increment and
decrement digits of a parameter value.
ENTER Accepts menu selection or data entry.
EXIT Cancels changed menu selection. Also
returns you to the previous menu level.
INFO Displays context-sensitive information
about the present menu level. Toggles
information message on/off.
cáÖìêÉ=OJP
Simplified model of measurement operation
Arm
Layer
Idle
Scan
Layer
Measure
Layer
Yes
No
Arm
Count
Another
Arm
?
Another
Scan
?
Scan
Count
Another
Measure-
ment
?
Measure
Count
Measure
Event
Arm
Event
Scan
Event
Yes
No
Yes
No
Device