User manual

Front Panel Operation
3-82 2001-900-01 Rev. K/ August 2010
Since burst mode turns off autozero, an autozero refresh is
required once every 24 hours (by changing functions,
resolution, or issuing an autozero bus command).
3.8.2 Configuring data storage
The data storage configuration menu is used for the
following operations:
To acquire a burst of readings at high speed.
To select the data types stored in the buffer.
To select the buffer control.
To clear the buffer of readings and statistics.
To specify the number of readings to store.
To specify the source of readings to be stored.
Perform the following steps to display the CONFIG DATA
STORE menu:
1. From the normal reading display, press the CONFIG
key, and then the STORE key. The following menu is
displayed:
CONFIG DATA STORE
BURST-MODE DATA-GROUP CONTROL
CLEAR-ALL COUNT FEED
2. Use the cursor keys ( and ) to highlight the appro-
priate item and press ENTER to select it.
q~ÄäÉ=PJPP
Burst mode sequence
Action Result Annunciators
BURST MODE ON
ENTER BURST:00100 READINGS
Use , , , , ENTER, EXIT, or INFO
ARM and AUTO off
ENTER 00100 READING BURST
Use TRIG to start; EXIT to abort
TRIG (burst readings acquired)
(post-processing of readings)
BURST:00100 READINGS
Processing rdg #xx of 1000
BURST:00100 READINGS
Storage complete; press RECALL
BURST:00100 READINGS
Use , , , , ENTER, EXIT, or INFO
ARM on
* on
ARM and * off
RECALL Rdg#+00000
...
Rdg#+00099
EXIT BURST:00100 READINGS
Use , , , , ENTER, EXIT, or INFO
(press ENTER to loop back to start)
EXIT BURST MODE ABORTED
Use CONFIG STORE to resume
(normal reading display)
ARM and AUTO on
Notes:
1. This table assumes the unit was set for autorange before enabling burst.
2. Multiple displays of buffered readings are available when burst data is recalled. (See paragraph 3.5.4.)