User manual

To connect the Model 2000 multimeter to the IEEE-488 bus, follow these steps:
1. Line up the cable connector with the connector located on the rear panel. The con-
nector is designed so that it will fit only one way. Figure 4-4 shows the location of the
IEEE-488 connector.
2. Tighten the screws securely, making sure not to over tighten them.
3. Connect any additional connectors from other instruments as required for your appli-
cation.
4. Make certain that the other end of the cable is properly connected to the controller.
Most controllers are equipped with an IEEE-488 style connector, but a few may re-
quire a different type of connecting cable. See your controllers instruction manual for
information about properly connecting to the IEEE-488 bus.
NOTE You can only have 15 devices connected to a IEEE-488 bus, including the control-
ler. The maximum cable length is either 20- meters or two meters times the num-
ber of devices, whichever is less. Not observing these limits may cause erratic bus
operation.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
RS232
1 3 5
2
4 6
VMC
EXT TRIG
FUSE LINE
250mAT
(SB)
100 VAC
120 VAC
125mAT
(SB)
220 VAC
240 VAC
MADE IN
U.S.A.
INPUT
500V
PEAK
350V
PEAK
1000V
PEAK
TRIGGER
LINK
SENSE
W 4W
HI
LO
!
LINE RATING
50, 60
400HZ
22 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
!
KEITHLEY
Figure 4-4
IEEE-488 con-
nector location
2000-900 (J - Aug 2010) BOOK.fm Page 11 Wednesday, October 12, 2011 12:29 PM