User manual
Choose one of three languages to program the Model 2000 multimeter:
• SCPI (Signal Oriented Measurement Commands)
• Keithley Models 196/199 Digital Multimeter
• Fluke Model 8840A/8842A Digital Multimeter
The factory sets the language selection as SCPI.
You only can select a programming language from the front panel. The language selec-
tion is stored in non-volatile memory, which means it does not change when power has been
off or after a remote interface reset.
Table 4-1 shows the languages supported by the two available interfaces:
As you make your language selection, keep in mind that the language you select deter-
mines the remote operations allowed.
To select a programming language, follow these steps:
1. Access the GPIB configuration options by pressing SHIFT then GPIB.
You see GPIB:ON with GPIB blinking.
2. Select the language configuration option by pressing the ENTER key twice.
You see: LANG:<name>
Move to the language selection field by pressing the key.
Select the programming language you want by pressing the or key until you see
the appropriate language.
The menu scrolls through these choices: SCPI, 199 (Keithley Models 196/199), and
8842 (Fluke Model 8840A/8842A).
5. Confirm your selection by pressing ENTER. The multimeter returns to the measure-
ment mode.
Standard Commands for Programmable Instruments (SCPI) is fully supported by the
GPIB and RS-232 interfaces. Always calibrate the Model 2000 multimeter using the SCPI
language.
Table 4-1
Language support
Language GPIB RS-232
SCPI
Keithley Models 196/199
Fluke Model 8840A/8842A
Yes
Yes
Yes
Yes
No
No