The Model 2000 has other front panel operations. Saving and restoring setup information
is described in Section 2 — Basic Measurements. Selecting the remote interface and lan-
guage is covered in Section 4 — Remote Operation.
Self-test
The TEST selections are used as diagnostic tools to isolate problems within the Model
2000. Information on using these test procedures is included in the optional Model 2000 Re
-
pair Manual.
Calibration
The CAL selections are used to view the calibration date and next due date, to perform
calibration, and to view the number of times calibration has been performed. Some of the
items are password-protected to prevent unintended changing of calibration constants.
To view the calibration dates, press SHIFT-CAL. Press ENTER at the DATES prompt. The
first date is the last time calibration was performed. The NDUE date is the calibration due
date.
Running calibration is password-protected. Refer to the Model 2000 Calibration Manual
for details.
To view the calibration count, press ENTER at the COUNT prompt.