User manual

Low level considerations
For sensitive measurements, external considerations beyond the Model 2000 affect the
accuracy. Effects not noticeable when working with higher voltages are significant in micro
-
volt signals. The Model 2000 reads only the signal received at its input; therefore, it is im-
portant that this signal be properly transmitted from the source. The following paragraphs
indicate factors that affect accuracy, including stray signal pick-up and thermal offsets.
Shielding
AC voltages that are extremely large compared with the DC signal to be measured may
produce an erroneous output. Therefore, to minimize AC interference, the circuit should be
shielded with the shield connected to the Model 2000 INPUT LO (particularly for low level
sources). Improper shielding can cause the Model 2000 to behave in one or more of the
following ways:
Unexpected offset voltages.
Inconsistent readings between ranges.
Sudden shifts in reading.
To minimize pick-up, keep the voltage source and the Model 2000 away from strong AC
magnetic sources. The voltage induced due to magnetic flux is proportional to the area of
the loop formed by the input leads. Therefore, minimize the loop area of the input leads and
connect each signal at only one point.
NOTE Shielded cables should be used for input circuits to avoid interference caused by
conducting RF.
2001 MULTIMETER
Caution: Maximum Input = 750V RMS, 1000V peak, 8 x 10
7
V•Hz
AC Voltage
Source
Model 2000
Input Impedence = 1MΩ and 100pF
SHIFT