User manual
Trigger mode T0
T1
T2
T3
T4
T5
T6
T7
Continuous on Talk
One-shot on Talk
Continuous on GET
One-shot on GET
Continuous on X
One-shot on X
Continuous on External Trigger
One-shot on External Trigger
Reading mode B0
B1
B2
Readings from A/D converter
Individual readings from data store
All readings from data store (buffer dump)
Data store size I0
In
Disable data store
Data store of n (n=1 to 500), fill and stop
Interval Q0
Qn
Default interval, 175msec (SELECT OFF)
n=interval in milliseconds (15msec to 999999msec)
Value V±nn.nnnn or
V±n.nnnnnnE+n
Zero value, simulated reference junction temperature
Default conditions L0
L1
Restore factory default conditions and save (L1)
Save present machine states as default conditions
Data format G0
G1
G2
G3
G4
G5
G6
G7
Reading with prefix.
Reading without prefix.
Reading and buffer location with prefix.
Reading and buffer location without prefix.
Reading and channel with prefix.
Reading and channel without prefix.
Reading, buffer location, and channel with prefix.
Reading, buffer location, and channel without prefix.
SRQ M0
M1
M2
M4
M8
M16
M32
Disable
Reading overflow
Data store full
Data store half full
Reading done
Ready
Error
EOI and bus hold-
off
K0
K1
K2
K3
Enable EOI and bus hold-off on X
Disable EOI, enable bus hold-off on X
Enable EOI, disable bus hold-off on X
Disable both EOI and bus hold-off on X
Terminator Y0
Y1
Y2
Y3
CR LF
LF CR
CR
LF
Table D-1 (cont.)
Models 196/199 device-dependent command summary
Mode Command Description
Models 196/199 and 8840A/8842A Commands D-3
2000-900 (J - Aug 2010) BOOK.fm Page 3 Wednesday, October 12, 2011 12:29 PM