:TIMer <n>
:TRIGger:[SEQuence[1]]:TIMer <n> Set interval for measure layer timer
Parameters <n> = 0.001 to 999999.999 Specify timer interval in seconds
Query :TIMer? Query programmed timer interval
Description These commands are used to set the interval for the timer. Note that the
timer is in effect only if the timer is the selected control source.
:SIGNal
:TRIGger[:SEQuence[1]]:SIGNal Bypass measure control source
Description This action command is used to bypass the specified control source when
you do not wish to wait for the programmed event. Keep in mind that the
instrument must be waiting for the appropriate event when the command
is sent. Otherwise, an error occurs and this command is ignored.
:SAMPle Command
:SAMPle:COUNt <NRf> Set sample count
Parameter <NRf> = 1 to 1024
Query :COUNt? Query the sample count
Description This command specifies the sample count. The sample count defines how
many times operation loops around in the trigger model to perform a de-
vice action.
NOTE If sample count is >1, you cannot use the :READ? command if
there are readings stored in the buffer.