User manual
:TRIGger commands
:COUNt <n>
:TRIGger[:SEQuence[1]]:COUNt <n> Set measure count
Parameters <n> = 1 to 9999 Specify count
INF Sets count to infinite
DEFault Sets count to 1
MINimum Sets count to 1
MAXimum Sets count to 9999
Query :COUNt? Queries programmed count
:COUNt? DEFault Queries *RST default count
:COUNt? MINimum Queries lowest allowable count
:COUNt? MAXimum Queries largest allowable count
Description This command is used to specify how many times operation loops around
in the trigger operation. For example, if the count is set to 10, operation
continues to loop around until 10 device actions are performed. After the
10th action, operation proceeds back up to the start of the trigger model.
Note that each loop places operation at the control source where it waits
for the programmed event.
2000-900 (J - Aug 2010) BOOK.fm Page 71 Wednesday, October 12, 2011 12:29 PM