User manual

:REAL:OFFSet <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:REAL:OFFSET <n>
Parameters <n> = -0.09999 to 0.09999Specify voltage offset at 0°C
DEFault 0.05463
MINimum -0.09999
MAXimum 0.09999
Query :OFFSet? Query voltage offset
:OFFSet? DEFault Query *RST default voltage offset
:OFFSet? MINimum Query lowest allowable voltage offset
:OFFSet? MAXimum Query largest allowable voltage offset
Description This command is used to specify the offset voltage at 0°C for the specified
reference junction.
:DIODe command
:RANGe[:UPPer] <NRf>
[:SENSe[1]]:DIODe:CURRent:RANGe[:UPPer] <NRf> Select current range for diode test
Parameters <NRf> = 0 to 1e-3 Specify diode test current
Query [UPPer]? Query selected range
Description There are three current ranges available for the diode test: 10µA range,
100µA range and the 1mA range. Range is selected by using this com-
mand to specify the expected current for the diode under test. The instru-
ment will then automatically select the appropriate range.
:CONTinuity command
:THReshold <n>
[SENSe[1]]:CONTinuity:THReshold <NRf> Specify threshold resistance
Parameters <NRf> = 1 to 1000 Specify threshold in ohms
Query :THReshold? Query threshold resistance
This command is used to specify the threshold resistance for the continu-
ity test. Continuity occurs when the measurement is less than or equal to
the specified threshold level.
2000-900 (J - Aug 2010) BOOK.fm Page 51 Wednesday, October 12, 2011 12:29 PM