User manual
Thermocouple commands
:TYPE <name>
[:SENSe[1]]:TEMPerature:TCouple:TYPE <name> Specify TC type
Parameters <name> = J Set operation for Type J thermocouples
K Set operation for Type K thermocouples
T Set operation for Type T thermocouples
Query :TYPE? Query thermocouple type
Description This command is used to configure the Model 2000 for the thermocouple
type that you are using to make temperature measurements.
These commands are used to configure the reference junction for thermo-
couple temperature measurements.
:RSELect <name>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:RSELect <name> Specify reference junction type.
Parameters <name> = SIMulated Use simulated temperature as reference
REAL Use a measured temperature as reference
Query :RSELect? Query reference junction type
Description This command is used to specify the type of reference junction that is go-
ing to be used for thermocouple temperature measurements. Specify
REAL if you are using an actual reference junction. The :REAL command
is then used to specify the desired reference temperature. Specify SIMu
-
lated if you wish to use a simulated reference temperature. The :SIMulat-
ed command is then used to specify the desired simulated reference
temperature.
2000-900 (J - Aug 2010) BOOK.fm Page 49 Wednesday, October 12, 2011 12:29 PM