:DIGits command
:DIGits <n>
[:SENSe[1]]:CURRent:AC:DIGits <n> Specify resolution for ACI
[:SENSe[1]]:CURRent:DC:DIGits <n> Specify resolution for DCI
[:SENSe[1]]:VOLTage:AC:DIGits <n> Specify resolution for ACV
[:SENSe[1]]:VOLTage:DC:DIGits <n> Specify resolution for DCV
[:SENSe[1]]:RESistance:DIGits <n> Specify resolution for Ω2
[:SENSe[1]]:FRESistance:DIGits <n> Specify resolution for Ω4
[:SENSe[1]]:PERiod:DIGits <n> Specify resolution for PER
[:SENSe[1]]:FREQuency:DIGits <n> Specify resolution for FREQ
[:SENSe[1]]:TEMPerature:DIGits <n> Specify resolution for TEMP
Parameters <n> = 4 3H digits
54H digits
65H digits
76H digits
DEFault 6H digits for DCI, DCV, Ω2, Ω4, FREQ, PER
5H digits for ACI, ACV, TEMP
MINimum 3H for DCI, ACI, ACV, DCV, Ω2, Ω4, TEMP
MAXIMUM 6H digits for DCI, ACI, ACV, DCV, Ω2, Ω4, TEMP
Query :DIGits? Query selected resolution
:DIGits? DEFault Query *RST default resolution
:DIGits? MINimum Query minimum allowable resoltuion
:DIGits? MAXimum Query maximum allowable resolution
Description These commands are used to select display resolution for the specified
measurement function.
Even though the parameters for this command are expressed as integers
(4 to 7), you can specify resolution using real numbers. For example, to
select 3
H digit resolution let <n> = 3.5, for 4H digit let <n> = 4.5, and so
on. Internally, the instrument rounds the entered parameter value to the
nearest integer.