User manual

:HOLD Command
The following commands are used to configure and control the Hold fea-
ture. For details on Hold, refer to “Trigger Model, Device Action” in this
section and “Hold” in Section 3.
:WINDow <NRf>
[:SENSe[1]]:HOLD:WINDow <NRf><name> Set Hold window
Parameter <NRf> = 0.01 to 20 Set window (percent)
Query :WINDow? Query Hold window.
Description This command is used to set the window for Hold. The window is ex-
pressed as a percent of the “seed” reading for the Hold process.
:COUNt <NRf>
[:SENSe[1]]:HOLD:COUNt <NRf> Specify Hold count.
Parameter <NRf> = 2 to 100 Specify Hold count
Query :COUNt? Query Hold count.
Description This command is used to specify the count for Hold. Count is the number
of readings that are compared to the “seed” reading during the Hold pro
-
cess.
:STATe <b>
[:SENSe[1]]:HOLD:STATe <b> Control (on/off) Hold
Parameters <b> = 0 or OFF Disable Hold
1 or ON Enable Hold
Query :STATe? Query state of Hold.
Description This command is used to enable or disable Hold. See “Hold” in Section 3
and “Trigger Model, Device Action” in this section for details on Hold.
2000-900 (J - Aug 2010) BOOK.fm Page 39 Wednesday, October 12, 2011 12:29 PM