The commands in this subsystem are used to control the display of the
Model 2000 and are summarized in Table 5-3.
:ENABle <b>
:DISPlay:ENABle <b> Control display circuitry
Parameters <b> = 0 or OFF Disable display circuitry
1 or ON Enable display circuitry
Query :ENABle? Query state of display
Description This command is used to enable and disable the front panel display cir-
cuitry. When disabled, the instrument operates at a higher speed. While
disabled, the display is frozen.
All front panel controls (except LOCAL) are disabled. Normal display op-
eration can be resumed by using the :ENABle command to enable the dis-
play or by putting the Model 2000 into local mode (press LOCAL).
:TEXT commands
:DATA <a>
:DISPlay[:WINDow[1]]:TEXT:DATA <a> Define message for display.
Parameter <a> = ASCII characters for message (maximum of 12 characters). The
characters must be enclosed in either double quotes (“ ”) or single
quotes (‘ ’).
Query :DATA? Query the defined text message.
Description These commands define the text message for display. A message can be
as long as 12 characters. A space counts as a character. Excess mes
-
sage characters results in an error.
2000-900 (J - Aug 2010) BOOK.fm Page 26 Wednesday, October 12, 2011 12:29 PM