User manual

Query :MUNits? Query units for mx+b
Description This command is used to specify the units data element for the mx+b cal-
culation. Use any three letters from ‘A’ through ‘Z’.
:PERCent <NRf>
:CALCulate [1]:KMATh:PERCent <NRf> Specify target value for percent calculation
Parameter <Nrf> = -1e8 to +1e8 Specify target value.
Query :PERCENt? Query percent target value
Description This command is used to specify the target value for the percent calcula-
tion.
:ACQuire
:CALCulate [1]:KMATh:PERCent:ACQuire Use input signal as target value
Description This action command is used to acquire the present input signal reading
and use it as the target value for the Percent calculation.
:STATe <b>
:CALCulate [1]:STATe <b> Control CALC1
Parameters <b> = 0 or off Disable CALC1 calculation
1 or on Enable CALC1 calculation
Query :STATe? Query state (on or off) of CALC1.
Description This command is used to enable or disable the CALC1 calculation. When
enabled, each instrument reading will reflect the selected calculation (see
:FORMat).
:DATA?
:CALCulate [1]:KMATh:DATA? Read CALC1 result
Description This query command is used to read the result of the CALC1 calculation.
If CALC1 is disabled or NONE is selected, the “raw” reading will be read.
:CALCulate2
These commands are used to configure and control the CALC2 opera-
tions on readings stored in the buffer.
2000-900 (J - Aug 2010) BOOK.fm Page 21 Wednesday, October 12, 2011 12:29 PM