User manual

Příkazy :RANGe
[:UPPer] <n>
[:SENSe[1]]:CURRent:AC:RANGe[:UPPer] <n> Set measurement range for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe[:UPPer] <n> Set measurement range for DCI
[:SENSe[1]]:VOLTage:AC:RANGe[:UPPer] <n> Set measurement range for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n> Set measurement range for DCV
[:SENSe[1]]:RESistance:RANGe[:UPPer] <n> Set measurement range for 2
[:SENSe[1]]:FRESistance:RANGe[:UPPer] <n> Set measurement range for 4
Parametry <n> = 0 až 3.1 Expected reading is amps (ACI and DCI)
0 to 757.5 Expected reading is AC volts (ACV)
0 to 1010 Expected reading in DC volts (DCV)
0 to 120e6 Expected reading is ohms (2 and 4)
DEFault 3 (ACI and DCI)
757.5 (ACV)
1000 (DCV)
100e6 and
MINimum 0 (All functions)
MAXimum Same as DEFault
Query :RANGe[:UPPer]? Query ACI measurement range
:RANGe[:UPPer]? DEFault Query *RST default range
:RANGe[:UPPer]? MINimum Query lowest measurement range
:RANGe[:UPPer]? MAXimum Query highest measurement range
Popis
Tento příkaz se používá k manuálnímu nastavení rozsahu měření pro specifikovanou funkci měření.
Rozsah se vybere určením očekávané hodnoty jako absolutní hodnoty. Model 2000 pak vybere
nejcitlivěí rozsah, který odpovídá očekávané hodnotě. Když například očekáváte hodnotu výsledku
přibližně 50 mV, jednoduše nechte parametr (<n>) = 0,05 (nebo 50e-3), aby vybral rozsah 100 mV.
:AUTO <b>
[:SENSe[1]]:CURRent:AC:RANGe:AUTO <b> Control auto range for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO <b> Control auto range for DCI
[:SENSe[1]]:VOLTage:AC:RANGe:AUTO <b> Control auto range for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO <b> Control auto range for DCV
[:SENSe[1]]:RESistance:RANGe:AUTO <b> Control auto range for 2
[:SENSe[1]]:FRESistance:RANGe:AUTO <b> Control auto range for 4
Parametry <b> = 1 or ON Enable auto range
0 or OFF Disable auto range
Query :AUTO? Query auto range (on or off)
Popis
Tyto příkazy se používají ke kontrole automatického rozsahu. Pokud je funkce povolena, přístroj
automaticky přejde do nejcitlivějšího rozsahu pro měření.
Příkaz automatického rozsahu (:RANGe:AUTO) se pojí s příkazem, který manuálně vybere rozsah
měření (:RANGe <n>). Když je funkce automatického rozsahu povolena, hodnota parametru pro
:RANGe <n> se změní na automaticky vybranou hodnotu rozsahu. Pokud funkce automatického
rozsahu není povolena, přístroj zůstane v automaticky zvoleném rozsahu a zadá-li se plat
příkaz :RANGe <n>, automaticky rozsah se vypne.
Příkazy :REFerence <n>
:REFerence <n>
[:SENSe[1]]:CURRent:AC:REFerence <n> Specify reference for ACI
[:SENSe[1]]:CURRent[:DC]:REFerence <n> Specify reference for DCI
[:SENSe[1]]:VOLTage:AC:REFerence <n> Specify reference for ACV
:SENSe[1]]:VOLTage[:DC]:REFerence <n> Specify reference for DCV
[:SENSe[1]]:RESistance:REFerence <n> Specify reference for 2
[:SENSe[1]]:FRESistance:REFerence <n> Specify reference for 4
[:SENSe[1]]:FREQuency:REFerence <n> Specify reference for FREQ
[:SENSe[1]]:PERiod:REFerence <n> Specify reference for PER
[:SENSe[1]]:TEMPerature:REFerence <n> Specify reference for TEMP
Parametry <n> = -3.1 to 3.1 Reference for ACI and DCI
-757.5 to 757.5 Reference for ACV
-1010 to 1010 Reference for DCV
0 to 120e6 Reference for 2 and 4
0 to 1.5e7 Reference for FREQ
0 to 1 Reference for PER
-200 to 1372 Reference for TEMP
DEFault 0 (all functions)
MINimum Minimum value for specified function
MAXimum Maximum value for specified function
Query :REFerence? Query programmed reference value
:REFerence? DEFault Query *RST default reference value
:REFerence? MINimum Query lowest allowable reference value
:REFerence? MAXimum Query largest allowable reference value
Popis
Tyto příkazy se používají k vytvoření referenční hodnoty pro specifikovanou funkci.
Je-li reference povolena, výsledkem bude algebraický rozdíl mezi vstupním signálem a referenč
hodnotou:
Výsledek = Vstupní signál – Referenč hodnota
Na čelním panelu se reference označuje jako REL.
Příkaz :REFerence <n> se pojí s příkazem :ACQuire. Poslední odeslaný příkaz (:REFerence <n>,
nebo :ACQuire) vytvoří referenci. Pokud se reference vytvoří pomocí příkazu :REFerence <n>,
vrací naprogramovanou hodnotu dotaz :REFerence?. Naopak platí, že když reference vytvoří
pomocí příkazu :ACQuire, dotazem :REFerence? Se vrací požadovaná referenční hodnota.
:STATe <b>
[:SENSe[1]]:CURRent:AC:REFerence:STATe <b> Control reference for ACI
[:SENSe[1]]:CURRent[:DC]:REFerence:STATe <b> Control reference for DCI
[:SENSe[1]]:VOLTage:AC:REFerence:STATe <b> Control reference for ACV
[:SENSe[1]]:VOLTage[:DC]:REFerence:STATe <b> Control reference for DCV
[:SENSe[1]]:RESistance:REFerence:STATe <b> Control reference for 2
[:SENSe[1]]:FRESistance:REFerence:STATe <b> Control reference for 4
[:SENSe[1]]:FREQuency:REFerence:STATe <b> Control reference for FREQ
[:SENSe[1]]:PERiod:REFerence:STATe <b> Control reference for PER
[:SENSe[1]]:TEMPerature:REFerence:STATe <b> Control reference for TEMP
Parametry <b> = 1 or ON Enable reference
0 or OFF Disable reference
Query :STATe? Query state of reference
Popis
Tyto příkazy se používají k povolení, nebo k zakázání reference pro určenou funkci.
Pokud je reference povolena, zobrazovaný výsledek bude obsahovat také naprogramovanou
referenční hodnotu.