DC OPERATING CHARACTERISTICS
2
Function Digits Readings/s PLCs
8
DCV (all ranges),
3, 4
5 10
DCI (all ranges), and
3, 7
30 1
Ohms (<10M range)
3, 5
50 1
3, 5
270 0.1
5
500 0.1
5
1000 0.04
5
2000 0.01
DC SYSTEM SPEEDS
2, 6
RANGE CHANGE
3
: 50/s.
FUNCTION CHANGE
3
: 45/s.
AUTORANGE TIME
3, 10
: <30ms.
ASCII READINGS TO RS-232 (19.2K BAUD): 55/s.
MAX. INTERNAL TRIGGER RATE: 2000/s.
MAX. EXTERNAL TRIGGER RATE: 400/s.
DC GENERAL
LINEARITY OF 10VDC RANGE: ±(1ppm of reading + 2ppm of range).
DCV, W, TEMPERATURE, CONTINUITY, DIODE TEST INPUT PROTECTION: 1000V, all ranges.
MAXIMUM 4WW LEAD RESISTANCE: 10% of range per lead for 100W and 1kW ranges; 1kW per
lead for all other ranges.
DC CURRENT INPUT PROTECTION: 3A, 250V fuse.
SHUNT RESISTOR: 0.1W for 3A, 1A, and 100mA ranges. 10W for 10mA range.
CONTINUITY THRESHOLD: Adjustable 1W to 1000W.
AUTOZERO OFF ERROR: Add ±(2ppm of range error + 5µV) for <10 minutes and ±1°C change.
OVERRANGE: 120% of range except on 1000V, 3A, and diode.
SPEED AND NOISE REJECTION
Rate Readings/s Digits
RMS Noise 10V
Range NMRR
12
CMRR
13
10 PLC 5 < 1.5 µV 60 dB 140 dB
1 PLC 50 < 4 µV 60 dB 140 dB
0.1 PLC 500 < 22 µV 80 dB
0.01 PLC 2000 < 150 µV 80 dB
DC NOTES
1. Add the following to “ppm of range” uncertainty:1V and 100V, 2ppm; 100mV, 15ppm; 100W, 15ppm; 1kW
<1MW, 2ppm; 10mA and 1A, 10ppm; 100mA, 40ppm.
2. Speeds are for 60Hz operation using factory default operating conditions (*RST). Autorange off, Display off,
Trigger delay = 0.
3. Speeds include measurement and binary data transfer out the GPIB.
4. Auto zero off.
5. Sample count = 1024, auto zero off.
6. Auto zero off, NPLC = 0.01.
7. Ohms = 24 readings/second.
8. 1 PLC = 16.67ms @ 60Hz, 20ms @ 50Hz/400Hz. The frequency is automatically determined at power up.
9. For signal levels >500V, add 0.02ppm/V uncertainty for the portion exceeding 500V.
10. Add 120ms for ohms.
11. Must have 10% matching of lead resistance in Input HI and LO.
12. For line frequency ±0.1%.
13. For 1kW unbalance in LO lead.
14. Relative to calibration accuracy.
15. Specifications are for 4-wire ohms. For 2-wire ohms, add 1W additional uncertainty.
16. For rear inputs, add the following to temperature coefficient “ppm of reading” uncertainty 10MW 95ppm,
100MW 900ppm. Operating environment specified for 0º to 50ºC and 50% RH at 35ºC.
2000
-Digit Multimeter
DC Characteristics
Conditions: MED (1 PLC)
1
or SLOW (10 PLC)
or MED (1 P
LC) with filter of 10
Accuracy: ±(ppm of reading + ppm of range)
(ppm = parts per million)
(e.g., 10ppm = 0.001%)
Function Range Resolution
Test Current
or
Burden Voltage
(±5%)
Input
Resistance
Temperature
Coefficient
0°–18°C and 28°–50°C
24 Hour
14
23°C ± 1°
90 Day
23°C ± 5°
1 Year
23°C ± 5°
Voltage
100.0000 mV 0.1 µV
> 10 GW
30 + 30 40 + 35 50 + 35 2 + 6
1.000000 V 1.0 µV
> 10 GW
15 + 6 25 + 7 30 + 7 2 + 1
10.00000 V 10 µV
> 10 GW
15 + 4 20 + 5 30 + 5 2 + 1
100.0000 V 100 µV
10 MW ±1%
15 + 6 30 + 6 45 + 6 5 + 1
1000.000 V
9
1 mV
10 MW ±1%
20 + 6 35 + 6 45 + 6 5 + 1
Resistance
15
100.0000
W
100
µW
1 mA 30 + 30 80 + 40 100 + 40 8 + 6
1.000000
kW
1
mW
1 mA 20 + 6 80 + 10 100 + 10 8 + 1
10.00000
kW
10
mW
100 µA 20 + 6 80 + 10 100 + 10 8 + 1
100.0000
kW
100
mW
10 µA 20 + 6 80 + 10 100 + 10 8 + 1
1.000000
MW
16
1
W
10 µA 20 + 6 80 + 10 100 + 10 8 + 1
10.00000
MW
11, 16
10
W
700
nA // 10MW
150 + 6 200 + 10 400 + 10 95 + 1
100.0000
MW
11, 16
100
W
700
nA // 10MW
800 + 30 1500 + 30 1500 + 30 900 + 1
Current
10.00000 mA 10 nA < 0.15 V 60 + 30 300 + 80 500 + 80 50 + 5
100.0000 mA 100 nA < 0.03 V 100 + 300 300 + 800 500 + 800 50 + 50
1.000000 A 1 µA < 0.3 V 200 + 30 500 + 80 800 + 80 50 + 5
3.00000 A 10 µA < 1 V 1000 + 15 1200 + 40 1200 + 40 50 + 5
Continuity 2W
1
kW
100
mW
1 mA 40 + 100 100 + 100 120 + 100 8 + 1
Diode Test
3.00000 V 10 µV 1 mA 20 + 6 30 + 7 40 + 7 8 + 1
10.00000 V 10 µV 100 µA 20 + 6 30 + 7 40 + 7 8 + 1
10.00000 V 10 µV 10 µA 20 + 6 30 + 7 40 + 7 8 + 1
Model 2000 specifications
Model 2000 specifications