Datasheet

1.888.KEITHLEY (U.S. only)
www.keithley.com
DIGITAL MULTIMETERS & SYSTEMS
A GREATER MEASURE OF CONFIDENCE
HIGH CREST FACTOR ADDITIONAL ERROR ±(% of reading)
7
CREST FACTOR: 1–2 2–3 34 4–5
ADDITIONAL ERROR: 0.05 0.15 0.30 0.40
AC OPERATING CHARACTERISTICS
2
Function Digits Readings/s Rate Bandwidth
ACV (all ranges), and
3
2s/reading SLOW 3 Hz–300 kHz
ACI (all ranges)
3
1.4 MED 30 Hz300 kHz
4
4.8 MED 30 Hz–300 kHz
3
2.2 FAST 300 Hz–300 kHz
4
35 FAST 300 Hz–300 kHz
ADDITIONAL LOW FREQUENCY ERRORS ±(% of reading)
Slow Med Fast
20 Hz – 30 Hz 0 0.3
30 Hz – 50 Hz 0 0
50 Hz – 100 Hz 0 0 1.0
100 Hz 200 Hz 0 0 0.18
200 Hz – 300 Hz 0 0 0.10
> 300 Hz 0 0 0
AC SYSTEM SPEEDS
2, 5
FUNCTION/RANGE CHANGE
6
: 4/s.
AUTORANGE TIME: <3s.
ASCII READINGS TO RS-232 (19.2K BAUD)
4
: 50/s.
MAX. INTERNAL TRIGGER RATE
4
: 300/s.
MAX. EXTERNAL TRIGGER RATE
4
: 300/s.
AC GENERAL
INPUT IMPEDANCE: 1MΩ ±2% paralleled by <100pF.
ACV INPUT PROTECTION: 1000Vp.
MAXIMUM DCV: 400V on any ACV range.
ACI INPUT PROTECTION: 3A, 250V fuse.
BURDEN VOLTAGE: 1A Range: <0.3V rms. 3A Range: <1V rms.
SHUNT RESISTOR: 0.1Ω on all ACI ranges.
AC CMRR: >70dB with 1kΩ in LO lead.
MAXIMUM CREST FACTOR: 5 at full scale.
VOLT HERTZ PRODUCT: 8 × 10
7
V·Hz.
OVERRANGE: 120% of range except on 750V and 3A ranges.
AC NOTES
1. Specifications are for SLOW rate and sinewave inputs >5% of range.
2. Speeds are for 60Hz operation using factory default operating conditions (*RST). Auto zero off, Auto range off,
Display off, includes measurement and binary data transfer out the GPIB.
3. 0.01% of step settling error. Trigger delay = 400ms.
4. Trigger delay = 0.
5. DETector:BANDwidth 300, NPLC = 0.01.
6. Maximum useful limit with trigger delay = 175ms.
7. Applies to non-sinewaves >5Hz and <500Hz (guaranteed by design for crest factors >4.3).
8. Applies to 0°–18°C and 28°–50°C.
9. For signal levels >2,2A, add additional 0.4% to “of reading” uncertainty.
10. Typical uncertainties. Typical represents two sigma or 95% of manufactured units measure <0.35% of reading
and three sigma or 99.7% measure <1.06% of reading.
2000
-Digit Multimeter
True RMS AC Voltage and Current Characteristics
Accuracy
1
: ±(% of reading + % of range), 23°C ±5 °C
Voltage
Range Resolution Calibration Cycle 3 Hz–10 Hz
10
10 Hz–20 kHz 20 kHz–50 kHz 50 kHz–100 kHz 100 kHz–300 kHz
100.0000 mV 0.1 µV
1.000000 V 1.0 µV 90 Days 0.35 + 0.03 0.05 + 0.03 0.11 + 0.05 0.60 + 0.08 4 + 0.5
10.00000 V 10 µV
100.0000 V 100 µV 1 Year 0.35 + 0.03 0.06 + 0.03 0.12 + 0.05 0.60 + 0.08 4 + 0.5
750.000 V 1 mV
Temperature
Coefficient/°C
8
0.035 + 0.003 0.005 + 0.003 0.006 + 0.005 0.01 + 0.006 0.03 + 0.01
Current
Range Resolution Calibration Cycle 3 Hz–10 Hz 10 Hz–3 kHz 3 kHz–5 kHz
1.000000 A 1 µA 90 Day/1 Year 0.30 + 0.04 0.10 + 0.04 0.14 + 0.04
3.00000 A
9
10 µA 90 Day/1 Year 0.35 + 0.06 0.15 + 0.06 0.18 + 0.06
Temperature
Coefficient/°C
8
0.035 + 0.006 0.015 + 0.006 0.015 + 0.006
Model 2000 specifications
Model 2000 specifications