Appendix B: Nominal Min/Max Flow Rates
Minimum flow rates are taken from the smallest inside diameters and maximum
flow rates are taken from the largest inside diameters of the syringes supplied by
the eleven most widely used syringe manufacturers.
Nominal Minimum/Maximum Flow Rates for Various Syringes (Actual Limits will
vary depending on manufacturer)
Syringe µl/hr µl/min ml/hr ml/min
Size Min. Max. Min. Max. Min. Max. Min. Max.10
0.5 µl 0.0004 0.79
1 µl 0.0008 1.58
2 µl 0.0015 3.1
5 µl 0.0037 7.93
10 µl 0.0073 15.829
25 µl 0.0183 39.756
50 µl 0.0365 79.519
100 µl 0.0731 159.46
250 µl 0.1813 23.751
500 µl
1000 µl 0.7281 95.418
1 ml 0.7828 102.580
2 ml 2.8493 373.430
2.5 ml 1.8156 237.950
3 ml 2.5691 336.710
5 ml 4.9824 653.010
10 ml 7.2024 15.733
20 ml 12.5360 27.384
30 ml 16.1310 35.236
50 ml 24.4201 53.346
100 ml 0.087 91.20
140 ml 0.1053 106.60