Operating the KDS Allegro Peristaltic Pump System
KDS Allegro Peristaltic Pump System Users Manual
phone 508.429.6809 fax 508.893.0160
e-mail info@kdscientific.com web www.kdscientific.com
KD Scientific Allego Peristaltic Pump System User’s Manual
Publication 5613-004-REV-A
The break in programs are installed on the pump.
They are also located on the CD that ships with the pump.
KD Scientific may change the programs so are shown here
for reference only.
Allegro I: “I_BREAK_IN”
1. Ramp minimum flow to 300 ml/min 2½ min.
2. Ramp 300 to 650 ml/min min.
3. Constant 700 ml/min 5 min.
4. Constant 800 ml/min 5 min.
Allegro II:II_BREAK_IN”
1. Stepped5 minimum to 10 µl/min 1 min.
2. Stepped5 10 to 50 µl/min 2 min.
3. Ramp 50 µl/min to 10 ml/min 3 min.
4. Ramp 10 to 200 ml/min 3 min.
5. Stepped15 200 to 800 ml/min 5 min.
6. Constant 850 ml/min 3 min.
7. Constant 1700 ml/min 3 min.
Allegro III: “III_BREAK_IN
1. Ramp minimum to 100 ml/min min
2. Ramp 100 to 250 ml/min min.
3. Stepped15 500 to 1000 ml/min min.
4. Stepped15 900 to 1500 ml/min min.
5. Constant 1700 ml/min 5 min.
6. Constant 1700 ml/min 5 min.
APPENDIX B: TUBE BREAK-IN PROGRAMS