Operating the KDS Allegro Peristaltic Pump System
KDS Allegro Peristaltic Pump System User’s Manual
phone 508.429.6809 fax 508.893.0160
e-mail info@kdscientific.com web www.kdscientific.com
KD Scientific Allego Peristaltic Pump System User’s Manual
Publication 5613-004-REV-A
The break in programs are installed on the pump.
They are also located on the CD that ships with the pump.
KD Scientific may change the programs so are shown here
for reference only.
Allegro I: “I_BREAK_IN”
1. Ramp → minimum flow to 300 ml/min 2½ min.
2. Ramp ← 300 to 650 ml/min 2½ min.
3. Constant→ 700 ml/min 5 min.
4. Constant← 800 ml/min 5 min.
Allegro II: “II_BREAK_IN”
1. Stepped→5 minimum to 10 µl/min 1 min.
2. Stepped←5 10 to 50 µl/min 2 min.
3. Ramp → 50 µl/min to 10 ml/min 3 min.
4. Ramp ← 10 to 200 ml/min 3 min.
5. Stepped→15 200 to 800 ml/min 5 min.
6. Constant→ 850 ml/min 3 min.
7. Constant← 1700 ml/min 3 min.
Allegro III: “III_BREAK_IN”
1. Ramp → minimum to 100 ml/min 2½ min
2. Ramp ← 100 to 250 ml/min 2½ min.
3. Stepped→15 500 to 1000 ml/min 2½ min.
4. Stepped←15 900 to 1500 ml/min 2½ min.
5. Constant→ 1700 ml/min 5 min.
6. Constant← 1700 ml/min 5 min.
APPENDIX B: TUBE BREAK-IN PROGRAMS